Scatter radiation and image contrast control: Rad Tech
Scatter radiation is radiation that changes direction after X-ray photons interact with matter, such as the patient's tissues.
When the X-ray tube directs a beam of photons toward the patient, several things can happen. Some X-rays interact with the patient and are scattered away from the image receptor, while others are absorbed within the patient's tissues.
X-rays also pass through the patient without interacting and are called transmitted X-rays. Other X-rays are scattered inside the patient but still reach the image receptor.
When the transmitted X-rays and scattered X-rays reach the image receptor, they make up the image forming X-rays. Transmitted X-rays provide information about differences in tissue attenuation, meaning the tissues absorb or weaken the X-ray beam in different ways.
Scattered X-rays add extra exposure that creates a fog-like background across the image that reduces image contrast. As a radiologic technologist, your goal is to produce a diagnostic quality image while following the Alara principle of keeping radiation exposure as low as reasonably achievable for the diagnostic task.
Controlling scatter radiation is an important part of that balance because it improves image contrast. So an important characteristic of image quality is image contrast, or the visible difference among light, dark, and gray areas of an image.
Another important characteristic is contrast resolution, which describes the ability to distinguish adjacent tissues with similar attenuation characteristics.
Most of the scatter that degrades radiographic contrast is produced through Compton interactions between an X-ray and a loosely bound outer shell electron.
These interactions change the direction of an X-ray photon and reduce its energy. When scattered photons reach the image receptor, they add unwanted exposure without providing useful information about the anatomy.
This reduces image contrast and can make subtle differences between tissues harder to distinguish. Now there are 3 major factors that affect the amount of scatter radiation reaching the image receptor kilovoltage peak or KVP, field size, and anatomy thickness.
First, KVP measures the X-ray beam's energy, and it affects both penetration and scatter reaching the receptor. In general, increasing KVP can increase scatter, even when milliaperesconds or MAS is reduced.
Higher KVP increases beam penetration and can allow the technologist to use less MAS which may reduce patient dose. Second, field size affects scatter production.
A larger radiation field irradiates more tissue, which produces more scatter. Appropriate collimation limits the beam to the anatomy required for the examination, reducing both scatter and unnecessary patient exposure.
Third, anatomy thickness affects scatter. Thicker anatomy provides more tissue for X-rays to interact with, so more scatter is produced.
Although anatomy thickness is usually not modifiable, appropriate collimation and compression, when clinically appropriate and safe, can help manage additional scatter.
Together, these factors help determine whether a grid or another scatter control method is needed. Scatter radiation can be controlled in different ways, including beam restriction, grids, and the air gap technique.
There's also software that can compensate for the effects of scatter after image acquisition, rather than physically removing photons during exposure.
Beam restriction is one of the simplest and most important ways to control scatter because a smaller field irradiates less tissue, produces less scatter, and reduces patient dose.
An aperture diaphragm is a simple beam restrictor with a fixed opening that limits the size and shape of the X-ray field, though it's no longer used.
Nowadays, a variable aperture collimator uses adjustable shutters along with a light field and mirror system to help the technologist visualize and adjust the size of the X-ray field before making the exposure.
When beam restriction alone isn't enough to control scatter, a radiographic grid can be used to absorb scattered photons before they reach the image receptor.
A grid is made out of thin lead strips separated by low attenuation interspace material. The grid is placed between the patient and the image receptor.
The lead strips absorb scattered photons, while many primary photons pass through the inner spaces. This reduces the scatter that reaches the receptor.
Grid construction varies by the height and thickness of the lead strips, the width of the inner spaces, and strip frequency.
The grid ratio is the height of the lead strips divided by the thickness of the inner space material. For example, let's say a grid has lead strips that are 2.8 millimeters high and the inner space material is 0.4 millimeters thick.
Since 2.8 divided by 0.4 equals 7, the grid ratio is 7 to 1. In general, a higher grid ratio removes more scatter and improves image contrast, but it also leaves less room for positioning error and usually requires more exposure to maintain adequate receptor exposure.
The Bucky factor compares the exposure needed with a grid to the exposure needed without one to maintain similar image receptor exposure.
A higher Bucky factor means a larger exposure is needed, which usually increases patient dose. For example, if an exposure of 20 milliaperesconds is needed with a grid and 5 milliaperesconds is needed without a grid to maintain similar receptor exposure, the Bucky factor is 20 divided by 5, which equals 4.
This means the grid requires 4 times more exposure than the non-grid technique. In general, grids with higher grid ratios remove more scatter, but they also tend to have higher bucky factors.
The exact bucky factor depends on the grid design and imaging conditions. Together, the grid ratio and Bucky factor summarize the grid trade-off.
Better contrast versus increased radiation exposure. Now there are some different grid designs.
A moving grid is moved during the exposure to blur grid lines that can be seen on the image. The movement makes grid lines less visible, but it doesn't remove the need for correct centering.
Another type of grid is a virtual grid. This type of software-based scatter correction is not a physical grid and does not stop scatter photons from reaching the detector.
Instead, it estimates the amount of scatter and processes the image to reduce its visual effects. It can be useful in mobile or bedside imaging, where physical grid alignment is difficult.
Grid cut-off occurs when a grid absorbs more primary radiation than intended because the X-ray beam and grid are misaligned.
The result is reduced receptor exposure, which may be more pronounced at the image edges. There are 4 main errors that can occur.
An off-level grid occurs when the tube, grid, or both are angled across the direction of the lead strips. This causes cutoff across the image.
An off center grid occurs when the central ray is not centered to a focused grid. The resulting loss of receptor exposure is usually fairly uniform across the image.
More than one grid error can occur at once. To prevent cutoff, check the grid label and orientation.
Center the central ray, keep the beam perpendicular to the grid, and avoid tube angulation across the lead strips. The major disadvantage of a physical radiographic grid is increased patient radiation dose because the grid absorbs some useful image forming photons along with scatter, exposure factors must generally be increased to maintain adequate image receptor exposure.
When automatic exposure control is used, this increase may occur as a longer exposure rather than a manual change in MAS.
As grid ratio increases, scatter cleanup improves, but patient dose also tends to increase. Grid selection also depends on part thickness.
The general rule of thumb is if the body part is thicker than 10 centimeters or 4 inches, then a grid should be used. Higher KVP examinations and thicker anatomy may require more scatter control, but grid choice should follow the examination protocol.
The goal is to balance scatter reduction, image contrast, and patient radiation dose. Use a grid when its expected diagnostic benefit justifies the exposure increase.
The air gap technique is an alternative to using a physical grid. The image receptor is moved away from the patient, increasing the object to image distance by about 10 to 15 centimeters.
Because scattered photons travel in many directions, fewer reach the image receptor, while primary photons continue forward.
Technical factors may need adjustment, but patient dose is not automatically higher. Removing a physical grid reduces the exposure needed, while the larger distance and loss of scatter may increase it.
The net effect depends on the imaging geometry and protocol. There's also an increased risk for magnification and geometric on sharpness.
Increasing the source to image distance and using an appropriate focal spot can help limit these effects. This technique is used for selected chest magnification and angiographic applications.
All right, as a quick recap. Scatter radiation changes direction after X-ray photons interact with matter, such as the patient's tissues.
Transmitted radiation passed through the patient without interacting. Transmitted X-rays carry useful anatomical information, while scattered X-rays add unwanted signal and reduce image contrast.
The three major factors that influence scatter are KVP, field size, and anatomy thickness. In general, larger fields and thicker anatomy produce more scatter, and higher KVP can increase the scatter contribution reaching the image receptor.
Radiologic technologists can control scatter through appropriate beam restriction, grid use, and the air gap technique. Careful collimation, appropriate grid use, and accurate grid positioning help produce diagnostic quality images, while keeping patient radiation exposure as low as reasonably achievable for the diagnostic task.
- "Radiologic science for technologists: Physics, biology, and protection (13th ed.)" Elsevier (2026)
- "Principles of radiographic imaging: An art and a science (7th ed.)" Elsevier (2026)