Radiographic exposure factors and technique fundamentals: Rad Tech
Exposure factors are the variables that radiographers use to determine the energy and intensity of the x-ray beam. As a radiologic technologist, choosing exposure factors appropriately helps produce a diagnostic image while managing patient radiation dose.
The four prime exposure factors are kilovolt peak, or kVp; x-ray tube current, measured in milliamperes, or mA; exposure time; and source-to-image receptor distance, or SID. Each factor has a different job.
kVp primarily controls beam energy and penetrability. Milliamperes control the rate of x-ray production, while exposure time controls how long x-rays are produced. Together, milliamperes and exposure time determine the total quantity of x-ray photons produced. SID is the distance from the X-ray tube’s focal spot to the image receptor.
Now, kVp is the primary control of x-ray beam energy. Increasing kVp produces higher-energy x-rays by accelerating electrons across the x-ray tube with greater kinetic energy. Because higher-energy x-rays pass through tissue more readily...
kVp also controls beam penetrability. For example, when acquiring an abdominal radiograph, a higher kVp is needed than when acquiring a radiograph of a hand because the abdomen is thicker and contains more tissue for the beam to penetrate.
The radiologic technologist’s goal is to select the appropriate kVp for the anatomy of interest, because higher kVp doesn't automatically mean better image quality. Although a higher kVP improves penetration and increases image receptor exposure, it also increases radiation scatter and reduces the contrast differences between tissues.
While kVp primarily controls beam energy, milliamperes and exposure time control beam quantity. Milliamperage determines how many electrons per second travel from the cathode to the anode. These electrons strike the anode target, where their kinetic energy is converted into heat and x-ray photons. When more electrons strike the target, more x-ray photons are produced.
Exposure time determines how long x-rays are produced.
Together, mA and exposure time determine milliampere-seconds, or mAs, which is the total quantity of x-ray photons produced during an exposure. When the other exposure factors remain unchanged, increasing mAs increases image receptor exposure.
Patient dose is more closely related to total mAs than to mA alone. When the other exposure factors remain unchanged, increasing mAs increases patient radiation dose. Too little mAs means fewer photons reach the image receptor, while a high mAs adds patient dose without necessarily improving image quality.
In clinical practice, mA and exposure time can be balanced. If motion is a concern, the radiographer can select a shorter exposure time and increase mA enough to maintain the required mAs. This preserves x-ray quantity and image receptor exposure while reducing the opportunity for motion blur.
The final prime exposure factor is source-to-image receptor distance, which is the distance from the x-ray tube’s focal spot to the image receptor. The focal spot is the area on the anode surface bombarded by electrons where x-rays are produced. SID affects the intensity of the x-ray beam at the image receptor, but the energy of the x-rays does not change.