Scatter radiation and image contrast control: Rad Tech

Ver video solo

Scatter radiation is radiation that changes direction after x-ray photons interact with matter, such as the patient’s tissues. When the x-ray tube directs a beam of photons toward the patient, several things can happen. Some x-rays interact with the patient and are scattered away from the image receptor, while others are absorbed within the patient’s tissues. X-rays also pass through the patient without interacting and are called transmitted x-rays. Other x-rays are scattered inside the patient but still reach the image receptor. When the transmitted x-rays and scattered x-rays reach the image receptor, they make up the image-forming x-rays.

Transmitted x-rays provide information about differences in tissue attenuation, meaning the tissues absorb or weaken the x-ray beam in different ways. Scattered x-rays add extra exposure that creates a fog-like background across the image that reduces image contrast.

As a radiologic technologist, your goal is to produce a diagnostic-quality image while following the ALARA principle of keeping radiation exposure as low as reasonably achievable for the diagnostic task. Controlling scatter radiation is an important part of that balance because it improves image contrast.

So, an important characteristic of image quality is image contrast, or the visible difference among light, dark, and gray areas of an image. Another important characteristic is contrast resolution, which describes the ability to distinguish adjacent tissues with similar attenuation characteristics.

Most of the scatter that degrades radiographic contrast is produced through Compton interactions between an x-ray and a loosely-bound outer-shell electron. These interactions change the direction of an x-ray photon and reduce its energy. When scattered photons reach the image receptor, they add unwanted exposure without providing useful information about the anatomy. This reduces image contrast and can make subtle differences between tissues harder to distinguish.

Now, there are three major factors that affect the amount of scatter radiation reaching the image receptor: kilovoltage peak, or kVp; field size; and anatomy thickness.

First, kVp measures the x-ray beam’s energy, and it affects both beam penetration and scatter reaching the receptor. In general, increasing kVp can increase scatter, even when milliampere-seconds, or mAs, is reduced. Higher kVp increases beam penetration and can allow the technologist to use less mAs, which may reduce patient dose.

Second, field size affects scatter production. A larger radiation field irradiates more tissue, which produces more scatter. Appropriate collimation limits the beam to the anatomy required for the examination, reducing both scatter and unnecessary patient exposure.

Third, anatomy thickness affects scatter. Thicker anatomy provides more tissue for x-rays to interact with, so more scatter is produced. Although anatomy thickness is usually not modifiable, appropriate collimation and compression, when clinically appropriate and safe, can help manage additional scatter. Together, these factors help determine whether a grid or another scatter-control method is needed.