X-ray emission: Rad Tech

Ver video solo

Producing a diagnostic radiograph requires control of the x-ray beam, specifically its intensity and energy. Intensity reflects photon quantity, whereas energy determines penetrability, or the beam's ability to pass through tissue. These characteristics affect image receptor exposure, image quality, and patient dose. As a radiologic technologist, understanding the factors that affect the x-ray beam guides the selection of exposure techniques, supports diagnostic image quality, and limits unnecessary radiation exposure.

Now, x-ray intensity is the quantity, or number, of x-ray photons in the useful beam. More photons increase image receptor exposure, while too few can increase image noise. Because photon quantity also affects patient dose, intensity must be controlled. The main factors that affect x-ray intensity are milliampere-seconds, or mAs, kilovolt peak or kVp, distance, and filtration.

A milliampere is the rate of electron flow from cathode to anode in the X-ray imaging system, so miliampere-seconds equals the number of electrons produced in a certain amount of time. mAs controls x-ray beam quantity, so beam intensity rises in direct proportion to mAs. When more electrons strike the target, more photons are produced, increasing intensity and overall image receptor exposure. Conversely, decreasing mAs lowers intensity and image receptor exposure. Because mAs changes photon quantity rather than energy, it does not change the penetrability of the x-ray beam, though it directly impacts patient radiation dose.

While mAs affects only quantity, kVp affects both x-ray beam quantity and energy. Remember that kVp is the maximum voltage applied across the x-ray tube during exposure, so increasing kVp gives electrons more kinetic energy as they cross the tube. This produces more photons with higher average and maximum energies. As a result, the beam becomes more intense and more penetrating, allowing more photons to reach the image receptor.

Now, distance changes intensity without changing photon energy. As the distance from the x-ray source increases, the beam spreads over a larger area, reducing the number of photons that strike each unit of area, and intensity falls rapidly. This relationship is described by the inverse square law and helps radiologic technologists adjust mAs to maintain image receptor exposure when the source-to-image receptor distance changes.